Vital Signs                 

 

 

Blood Pressure

bulletBlood Pressure Technique: Oscillometric measurement of systolic, diastolic and mean arterial pressure and pulse rate.
bulletCuff Inflation Adjustment Range: Adult/Pediatric: 100 to 250 mmHg; 5 mmHg increments Neonate: 100 to 140 mmHg; 5 mmHg increments
bulletCuff Pressure Range: Adult/Pediatric: 0 to 290 mmHg Neonate: 0 to 140 mmHg Automatically released if error condition is detected, power fails, or unit is switched off.
bulletTarget Inflation Pressure Range: User-selectable cuff inflation pressure Adult/Pediatric: nominally 160 mmHG Neonate: nominally 110 mmHg
bulletDetermination Time: 20 seconds to 45 seconds typical Adult - 120 secs maximum Neonate - 85 secs maximum Accuracy: Meets or exceeds SP 10-1992 AAMI standards
bulletPressure Preset Levels: Resolutions of 5 mmHg up to Maximum Adult/Neonate limits
bulletSystolic Determinations: Adult: Max: 245 mmHg; Min: 30 mmHg Neonate: Max: 140 mmHg; Min: 40 mmHg
bulletDiastolic Determinations: Adult: Max: 195 mmHg; Min: 10 mmHg Neonate: Max: 100 mmHg; Min: 20 mmHg
bulletMAP Determinations: Adult: Max: 215 mmHg; Min: 15 mmHg Neonate: Max: 115 mmHg; Min: 30 mmHg
bulletOperating Modes: AUTOMATIC, STAT, MANUAL Modes. Press START/STOP key to cancel.
bulletAutomatic Mode (Blood Pressure) Cycle Time: Preset at 3 minutes. User can select blood pressure determination frequency in varied increments between 1 and 120 minutes or operate by manual intervention. The result of the most recent determination will be displayed for up to 90 minutes (AUTO) or 2 minutes (MANUAL).
bulletStat Mode: Rapid BP determination with early systolic display within 5-minute period. Pulse Rate Range: Adult/Pediatric: 30 - 200 BPM Neonate: 30 200 BPM
bulletPulse Rate Accuracy: BP±3.5%
bulletOverpressure Cut-off: Adult/Pediatric: 300 - 330 mmHg Neonate: 150 - 165 mmHg

Alarms

bulletSystem Alarm Functions: Brief alarm test at power turn-on, alarm for no determination and a continuous alarm for microprocessor failure. Audible alarms with diagnostic codes displayed in pulse rate window for system failure.
bulletPatient Alarm Functions: Systolic: Preset at 30 mmHg low and 180 mmHg high Diastolic: Preset at 15 mmHg low and 130 mmHg high
bulletMAP: Preset at 50 mmHg low and 140 mmHg high for adult/pediatric and neonate
bulletPulse Rate: Preset at 40 BPM low and 160 BPM high
bulletSpO2 Alarm Limits: Preset at off (100%) high and 90% low. Range: High 51-100% (100% = off) Low 50-99%
bulletAlarm Indicators: LED with audio alarm silence key. LED only flashes if audio alarm silenced and an alarm condition exists. Diagnostic data indicated on LCD.

Trend

bulletPrior Data: Monitor stores 24 hours of data; up to 100 readings within time period.

Environmental

bulletOperating Temperatures: +41°F to +104°F (+5°C to +40°C)
bulletAtmospheric Pressure: 700 to 1060 hPa
bulletStorage Temperature: -4°F to +122°F (-20°C to +50°C)
bulletHumidity: 0 to 95% noncondensing

 

 

 

Power Requirements

bulletInput Power: AC input voltage: 100-240 VAC, 50/60 Hz (nominal) 90-253 VAC, 47-63 Hz (range), 50VA. DC input voltage: 24 VDC, 36VA, supplied from a source conforming to IEC 601-1.
bulletFuse: AC input is protected by two internal fuses. DC input line is protected by an internal auto-resetting fuse. Battery: 12 volt, 2.3 amp-hours protected by internal auto-resetting fuse.
bulletBattery: Sealed lead acid 12 volt 2.3 amp-hour, minimum operation time of 2 hours (5-minute auto cycle with adult cuff, temperature, SpO2 continuous and printer activated. Longer battery life with vital signs at less frequent intervals). Time for full recharge is 2 hours from full discharge with monitor switched off.
bulletLow Battery Warning: Flashing LED and audio tone alert user of low battery condition.
bulletPower Cable: Domestic: 10-ft, detachable, gray-jacketed, 16-gauge terminated with 3-prong hospital grade plug. International: 7.5-ft (2.3m), detachable, black jacketed, 16-gauge terminated with appropriate plug.

**IVAC® Temperature (PRO 200 & PRO 400 only)

bulletScale: °Fahrenheit(F) °Celsius(C)
bulletRangePredictive mode: Max: 42.2° C; 108.0° Min: 31.6° C; 88.9° F
bulletMonitor Mode: Max: 42.2° C; 108.0° F Min: 26.7° C; 80.0° F
bulletMonitor Mode Accuracy: ±0.1°C ±0.2°F (when tested in a calibrated liquid bath; meets ASTM E1112, Table 1, in range specified)
bulletPredictive Mode Accuracy: ±1.0°F ±0.6°C Determination time less than 60 seconds
Use only IVAC probes and probe covers. The size, shape, and thermal characteristics of the probe covers can affect the performance of the instrument. Inaccurate readings or retention problems may occur unless IVAC probes and probe covers are used. **IVAC is a trademark of Alaris Medical, Inc.

*NELLCOR™ Pulse Oximetry (SpO2) (PRO 300 & PRO 400 only)

bulletAccuracy (± 1 Standard Deviation): Adult/neonate 70 - 100% ±3.0 digits; under 70% unspecified
bulletSaturation Range: 0 - 100%
bulletPulse Rate Range: 20 - 250 BPM Pulse Rate Accuracy: SpO2, ±3 BPM
bulletSaturation Pitch: Pitch changes with saturation. Volume selectable from off to 9.
*NELLCOR is a trademark of Mallinckrodt, Inc.

Physical Dimensions

bulletDimensions: 9.8 in (H) x 9.8 in (W) x 6.9 in (D) 24.8 cm (H) x 24.8 cm (W) x 17.5 cm (D)
bulletWeight: 7.8 pounds; 3.5 kg (including battery)
bulletMountings: Self-supporting on rubber feet. Pole mountable. Carrying handle recessed into top of monitor.
bulletPortability: Can be hand-carried by the recessed handle. When attached to the 3215 Rolling Stand, the monitor and accessories can be wheeled from patient to patient.

Printer

bulletPaper Width: 1.75 inches
bulletWriting Method: Direct thermal
bulletPaper Environmental requirements: Temperature: 100°F (38°C) maximum Humidity: 80% maximum

 

 