Blood Pressure
 | Blood Pressure Technique: Oscillometric
measurement of systolic, diastolic and mean arterial pressure and
pulse rate.
 | Cuff Inflation Adjustment Range:
Adult/Pediatric: 100 to 250 mmHg; 5 mmHg increments Neonate: 100 to
140 mmHg; 5 mmHg increments
 | Cuff Pressure Range: Adult/Pediatric: 0 to 290
mmHg Neonate: 0 to 140 mmHg Automatically released if error condition
is detected, power fails, or unit is switched off.
 | Target Inflation Pressure Range:
User-selectable cuff inflation pressure Adult/Pediatric: nominally 160
mmHG Neonate: nominally 110 mmHg
 | Determination Time: 20 seconds to 45 seconds
typical Adult - 120 secs maximum Neonate - 85 secs maximum Accuracy:
Meets or exceeds SP 10-1992 AAMI standards
 | Pressure Preset Levels: Resolutions of 5 mmHg
up to Maximum Adult/Neonate limits
 | Systolic Determinations: Adult: Max: 245 mmHg;
Min: 30 mmHg Neonate: Max: 140 mmHg; Min: 40 mmHg
 | Diastolic Determinations: Adult: Max: 195
mmHg; Min: 10 mmHg Neonate: Max: 100 mmHg; Min: 20 mmHg
 | MAP Determinations: Adult: Max: 215 mmHg; Min:
15 mmHg Neonate: Max: 115 mmHg; Min: 30 mmHg
 | Operating Modes: AUTOMATIC, STAT, MANUAL
Modes. Press START/STOP key to cancel.
 | Automatic Mode (Blood Pressure) Cycle Time:
Preset at 3 minutes. User can select blood pressure determination
frequency in varied increments between 1 and 120 minutes or operate by
manual intervention. The result of the most recent determination will
be displayed for up to 90 minutes (AUTO) or 2 minutes (MANUAL).
 | Stat Mode: Rapid BP determination with early
systolic display within 5-minute period. Pulse Rate Range:
Adult/Pediatric: 30 - 200 BPM Neonate: 30 200 BPM
 | Pulse Rate Accuracy: BP±3.5%
 | Overpressure Cut-off: Adult/Pediatric: 300 -
330 mmHg Neonate: 150 - 165 mmHg |
| | | | | | | | | | | | |
Alarms
 | System Alarm Functions: Brief alarm test at
power turn-on, alarm for no determination and a continuous alarm for
microprocessor failure. Audible alarms with diagnostic codes displayed
in pulse rate window for system failure.
 | Patient Alarm Functions: Systolic: Preset at
30 mmHg low and 180 mmHg high Diastolic: Preset at 15 mmHg low and 130
mmHg high
 | MAP: Preset at 50 mmHg low and 140 mmHg high
for adult/pediatric and neonate
 | Pulse Rate: Preset at 40 BPM low and 160 BPM
high
 | SpO2 Alarm Limits: Preset at off (100%) high
and 90% low. Range: High 51-100% (100% = off) Low 50-99%
 | Alarm Indicators: LED with audio alarm silence
key. LED only flashes if audio alarm silenced and an alarm condition
exists. Diagnostic data indicated on LCD. |
| | | | |
Trend
 | Prior Data: Monitor stores 24 hours of data;
up to 100 readings within time period. |
Environmental
 | Operating Temperatures: +41°F to +104°F (+5°C
to +40°C)
 | Atmospheric Pressure: 700 to 1060 hPa
 | Storage Temperature: -4°F to +122°F (-20°C
to +50°C)
 | Humidity: 0 to 95% noncondensing |
| | |
|
Power Requirements
 | Input Power: AC input voltage: 100-240 VAC,
50/60 Hz (nominal) 90-253 VAC, 47-63 Hz (range), 50VA. DC input
voltage: 24 VDC, 36VA, supplied from a source conforming to IEC 601-1.
 | Fuse: AC input is protected by two internal
fuses. DC input line is protected by an internal auto-resetting fuse.
Battery: 12 volt, 2.3 amp-hours protected by internal auto-resetting
fuse.
 | Battery: Sealed lead acid 12 volt 2.3
amp-hour, minimum operation time of 2 hours (5-minute auto cycle with
adult cuff, temperature, SpO2 continuous and printer activated. Longer
battery life with vital signs at less frequent intervals). Time for
full recharge is 2 hours from full discharge with monitor switched
off.
 | Low Battery Warning: Flashing LED and audio
tone alert user of low battery condition.
 | Power Cable: Domestic: 10-ft, detachable,
gray-jacketed, 16-gauge terminated with 3-prong hospital grade plug.
International: 7.5-ft (2.3m), detachable, black jacketed, 16-gauge
terminated with appropriate plug. |
| | | |
**IVAC® Temperature (PRO 200 & PRO 400 only)
 | Scale: °Fahrenheit(F) °Celsius(C)
 | RangePredictive mode: Max: 42.2° C; 108.0°
Min: 31.6° C; 88.9° F
 | Monitor Mode: Max: 42.2° C; 108.0° F Min:
26.7° C; 80.0° F
 | Monitor Mode Accuracy: ±0.1°C ±0.2°F (when
tested in a calibrated liquid bath; meets ASTM E1112, Table 1, in
range specified)
 | Predictive Mode Accuracy: ±1.0°F ±0.6°C
Determination time less than 60 seconds |
| | | |
Use only IVAC probes and probe covers. The size, shape, and
thermal characteristics of the probe covers can affect the performance of
the instrument. Inaccurate readings or retention problems may occur unless
IVAC probes and probe covers are used. **IVAC is a trademark of Alaris
Medical, Inc.
*NELLCOR™ Pulse Oximetry (SpO2) (PRO 300 & PRO
400 only)
 | Accuracy (± 1 Standard Deviation):
Adult/neonate 70 - 100% ±3.0 digits; under 70% unspecified
 | Saturation Range: 0 - 100%
 | Pulse Rate Range: 20 - 250 BPM Pulse Rate
Accuracy: SpO2, ±3 BPM
 | Saturation Pitch: Pitch changes with
saturation. Volume selectable from off to 9.
*NELLCOR is a trademark of Mallinckrodt, Inc.
Physical Dimensions
 | Dimensions: 9.8 in (H) x 9.8 in (W) x 6.9
in (D) 24.8 cm (H) x 24.8 cm (W) x 17.5 cm (D)
 | Weight: 7.8 pounds; 3.5 kg (including
battery)
 | Mountings: Self-supporting on rubber feet.
Pole mountable. Carrying handle recessed into top of monitor.
 | Portability: Can be hand-carried by the
recessed handle. When attached to the 3215 Rolling Stand, the
monitor and accessories can be wheeled from patient to patient. |
| | |
Printer
 | Paper Width: 1.75 inches
 | Writing Method: Direct thermal
 | Paper Environmental requirements:
Temperature: 100°F (38°C) maximum Humidity: 80% maximum |
| |
|
| | |
|
|